Remote control commands
R&S
®
RTO6
2120
User Manual 1801.6687.02 ─ 05
Usage:
Query only
BUS<m>:TBTO:RESult:FRAMe<n>:BITRate?
Returns the primary bit rate.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n>
*
Selects the frame.
Return values:
<PrimaryBitRate>
Range:
0 to 100000000000
Increment: 1
*RST:
0
Default unit: bps
Usage:
Query only
BUS<m>:TBTO:RESult:FRAMe<n>:CRC?
Returns the CRC (frame check) checksum of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n>
*
Selects the frame.
Return values:
<CRC>
Range:
0 to 4294967295
Increment: 1
*RST:
0
Usage:
Query only
BUS<m>:TBTO:RESult:FRAMe<n>:DATA?
Returns the data for the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n>
*
Selects the frame.
Return values:
<Data>
Usage:
Query only
Protocols