Remote control commands
R&S
®
RTO6
1731
User Manual 1801.6687.02 ─ 05
BUS<m>:UART:SBIT
<StopBits>
Sets the number of stop bits: 1; 1.5 or 2 stop bits are possible.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<StopBits>
B1 | B15 | B2
*RST:
B1
Usage:
Asynchronous command
BUS<m>:UART:SSIZe
<DataBits>
Sets the number of data bits in a message.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataBits>
Number of data bits. 9 data bits are only possible with parity =
none.
Range:
5 to 9
Increment: 1
*RST:
8
Usage:
Asynchronous command
BUS<m>:UART:PACKets
<Packets>
Defines the method of packet separation. A packet is a number of subsequent words in
a date stream.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Packets>
NONE | EWORd | TOUT
NONE
Packets are not considered.
EWORd
End word, the end condition of a packet is a pattern. To define
the end word, use
TOUT
Defines a timeout between the packets. To set the timeout, use
*RST:
NONE
Protocols