User Manual 1801.6687.02 ─ 05
"Data (min/
max)"
Enter the bytes in msb first bit order. The maximum pattern length is 8
bytes.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
Remote command:
Symbol
Triggers on a symbol or on a wakeup pattern.
Trigger type: "Symbol"
"CAS/MTS"
Collision Avoidance Symbol / Media access Test Symbol. These sym-
bols are identical and can be sent in the optional symbol window at
the end of a communication cycle. They are used to avoid collisions
during the system start.
"Wakeup Pat-
tern"
The wakeup pattern is sent to activate the nodes of the system.
Remote command:
Error conditions
Triggers on one or more errors in the frame.
Trigger type: "Error conditions"
"FSS"
Error in a Frame Start Sequence. FSS follows the Transmission Start
Sequence TSS at the beginning of each frame.
"BSS"
Error in a Byte Start Sequence. The BSS is transmitted before each
byte.
"FES"
Error in Frame End Sequence. FES indicates the end of each frame.
"Header CRC"
Error in a cyclic redundancy check code of the header data which
covers mainly frame ID and payload length.
"Payload CRC" Error in a cyclic redundancy check code of the complete frame.
Remote command:
13.9.2.2
Triggering on FlexRay signals
Prerequisites: An FlexRay bus is configured, see .
1. Press the [Bus] key and selet the "Shortcuts" tab.
FlexRay (option R&S
RTO6-K520)