Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1170
User Manual 1801.6687.02 ─ 05
The used resources are listed in the "Details" tab. See also:
The" Turn on time" is measured as the time between the trigger point ("Trigger level
on" value is reached) and the time the given percentage of the "Steady state level" is
reached.
The" Turn off time" is measured as the time between the trigger point, delayed with the
set "Time" ("Trigger level on" value is reached) and the time the given percentage of
the "Steady state level" is reached, see
See also
and
The following remote commands are used for handling the measurement results:
●
●
●
19.5.2.2
Configuring turn on/ off
For details of the configuration settings, see
Chapter 19.5.2.3, "Turn on/off settings"
1. Select "Analysis">"Power".
2. Under "Power Path", select "Turn on/off ".
3. Connect the probes to the DUT as shown in the "Channels" tab:
Power path