Remote control commands
R&S
®
RTO6
2287
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..3
Parameters:
<CRCOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
command.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
and
*RST:
EQUal
TRIGger<m>:MPHY:CRC:MIN
<CRCMin>
Specifies the CRC bit pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<CRCMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CRC:MAX
<CRCMax>
Sets the end value of the CRC range if
is set to
INRange
or
OORange
.
Suffix:
<m>
.
1..3
Parameters:
<CRCMax>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MPHY:CREQ
<REQ>
Specifies the CREQ bit pattern.
Suffix:
<m>
.
1..3
Parameters:
<REQ>
ONE | ZERO | DC
Protocols