Mixed signal option (MSO, R&S
RTO6-B1)
R&S
®
RTO6
995
User Manual 1801.6687.02 ─ 05
You can also drag the bus waveforms on the display and scale them.
●
Show the result box of the decoded clocked bus signal
Each parallel bus is shown in a separate diagram, and the diagrams can be minimized
and arranged as usual.
The signal icon indicates the activities on the digital channels even if they are not dis-
played in the diagram, of if the acquisition has been stopped:
●
Blue: channel is low
●
Green: channel is high
●
Gray: channel state is changing
The display update rate of the oscilloscope is adapted to the visual perception of
human eyes, and it is slower than the acquisition rate. All analog and digital waveforms
that are acquired during one display update cycle are overlapped and displayed at
once. Thus you can see the cumulative occurrence of binary states and edge transi-
tions on the screen at once. Bus signals are not overlapped.
The trigger point is always visible on the display, it cannot be moved outside ("Restrict
horizontal position to acquisition range" is enabled automatically).
If digital channels are active, the trigger point is always visible on the display, it cannot
be moved outside.
To access and analyze one or more specific acquisitions, you can use the History
Viewer in the common way.
Furthermore, you can zoom in digital signals and bus signal in the same way as in ana-
log waveforms.
See also:
●
14.2.1
Parallel bus - decode table
Decoding is available for clocked parallel buses.
The decode table shows the decoded data words of the bus signal and the corre-
sponding time. Each clock edge corresponds to one row in the table. Below the table,
you can select the data format of the bus values.
Display