Remote control commands
R&S
®
RTO6
1627
User Manual 1801.6687.02 ─ 05
WITHin | OUTSide
Triggers on pulses inside or outside a given range. The range is
defined by the width ±delta.
SHORter | LONGer
Triggers on pulses shorter or longer than the given width.
*RST:
WITHin
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]
<SearchName>,<Width>
SEARch:TRIGger:PATTern:WIDTh[:WIDTh]?
<SearchName>
For the ranges WITHin and OUTSide, the width defines the center of a range that is
defined by the limits ±delta.
For the ranges SHORter and LONGer, the width defines the maximum and minimum
pulse width, respectively.
SEARch:TRIGger:PATTern:WIDTh:RANGe
. To set the
SEARch:TRIGger:PATTern:WIDTh:DELTa
.
Parameters:
<Width>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
5E-9
Default unit: s
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
SEARch:TRIGger:PATTern:WIDTh:DELTa
<SearchName>,<WidthDelta>
SEARch:TRIGger:PATTern:WIDTh:DELTa?
<SearchName>
Defines a range around the width value specified using
.
Parameters:
<WidthDelta>
Range:
0 to 432
Increment: 500E-12
*RST:
0
Default unit: s
Parameters for setting and query:
<SearchName>
String with the name of the search
Usage:
Asynchronous command
Search