Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1147
User Manual 1801.6687.02 ─ 05
To measure and display the current harmonic, the instrument uses the following mea-
surements and waveforms:
●
"MG 6" to measure the power waveform
●
"MG 7" to measure the spectrum voltage
●
"MG 8" to measure the spectrum current
●
"M2" Math 2 to calculate the power
●
"M3" Math 3 to calculate the FFT of the voltage
●
"M4" Math 4 to calculate the FFT of the current
The used resources are listed in the "Details" tab. See also:
The current harmonic results are defined as follows:
Result Table
Bar Chart Match
Description
Harmonic Index
Value of the X-Axis
The harmonic order
Frequency
-
The frequency value of the signal
Value
Value of the Y-Axis. Shown
by a green bar
The present value of the current harmonic
Maximum
Shown by a darkened green
bar
The maximum measured value
Standard limit
Shown by a white bar
The maxim allowed value according to the
selected standard
"Y max"
"Y max"
Sets the upper limit for the display of the Y
scale.
This value can be reset with the "Reset Max"
button
The following remote commands are used for handling the measurement results:
●
POWer:HARMonics:RESult<m>:FREQuency<n>:VALue?
●
POWer:HARMonics:RESult<m>:MAXValue<n>:VALue?
●
POWer:HARMonics:RESult<m>:STDinuse?
●
POWer:HARMonics:RESult<m>:STDValue<n>:VALue?
●
POWer:HARMonics:RESult<m>:VALue<n>:VALue?
●
19.3.3.2
Configuring current harmonic
For details of the configuration settings, see
Chapter 19.3.3.3, "Current harmonic set-
1. Select "Analysis">"Power".
2. Under "Power Analysis", select "Current Harmonic".
3. Connect the differential voltage probe and the current probe to the oscilloscope.
Input