Remote control commands
R&S
®
RTO6
2162
User Manual 1801.6687.02 ─ 05
TRIGger<m>:USB:WPID
<WithPIDCheck>
Defines, whether the packet ID error check shall be activated or not.
Parameters:
<WithPIDCheck>
ON | OFF
*RST:
OFF
TRIGger<m>:USB:WPOR
<WithPortCheck>
Defines, whether the port check shall be activated or not.
Parameters:
<WithPortCheck>
ON | OFF
*RST:
OFF
TRIGger<m>:USB:WSEU
<WithSEUCheck>
Defines, whether the SEU check shall be activated or not.
Parameters:
<WithSEUCheck>
ON | OFF
*RST:
OFF
TRIGger<m>:USB:WSTC
<WithSCCheck>
Defines, whether the Start / Complete (SC) check shall be activated or not.
Parameters:
<WithSCCheck>
ON | OFF
*RST:
OFF
24.17.21.3
Decode results
In all
BUS<m>:USB:PACKet<n>
commands, the suffix <m> selects the serial bus and
the suffix <n> selects the packet number in the decode table.
► Query the status of packet #19:
BUS:USB:PACKet19:STAT?
The result of this remote command query should be "OK".
....................................................................................2163
...........................................................................2164
.................................................................................. 2164
.................................................................................2165
BUS<m>:USB:PACKet<n>:ENDPoint?
...........................................................................2165
.....................................................................................2165
.............................................................................. 2165
Protocols