Acquisition and waveform setup
R&S
®
RTO6
189
User Manual 1801.6687.02 ─ 05
2. Tap "Fast segmentation" to activate the fast segmentation mode.
3. If you want to sample the maximum number of acquisitions in a series, enable
"Acquire maximum".
If you want to capture a defined number of acquisitions, disable "Acquire maxi-
mum" and enter the "Required" number of acquisitions.
4. Set the "Display time".
5.9.8
Using digital filters
Before using digital filters, you determine if you want to filter input channels only or if
the trigger signal is filtered too. The filter settings depend on this decision.
For details on filter settings and dependencies, see
Chapter 5.5.5, "Digital filter setup"
To filter the input channels only
1. Open "Menu" > "Vertical" > "Bandwidth" tab > "Digital Filter".
2. Set the "Trigger BW limit" to "Off".
3. Enter the "Channel BW limit" frequency for each filter.
4. Enable "Channel BW limit (On/Off)" for each channel to be filtered.
To filter the trigger signal
1. Open "Menu" > "Vertical" > "Bandwidth" tab > "Digital Filter".
2. Set the "Trigger BW limit" to "RF Reject".
3. Set the frequency limit for the filter: "Channel BW limit".
4. To filter the input channels too, enable "Channel BW limit (On/Off)" for each chan-
nel to be filtered.
The trigger filter settings are applied also to these input channels.
Setting up the waveform