Remote control commands
R&S
®
RTO6
1308
User Manual 1801.6687.02 ─ 05
ABS
Position in time and voltage values, or in other units depending
on the waveform character. Absolute positions move with the
waveform display when the scales, the vertical position or offset,
or the reference point are changed.
Use
DISPlay:SIGNal:LABel:HORizontal:ABSolute:
DISPlay:SIGNal:LABel:HORizontal:
REL
Fixed label position in percent of the screen counting from the
upper left corner.
Use
DISPlay:SIGNal:LABel:HORizontal:RELative:
DISPlay:SIGNal:LABel:VERTical:
Parameters for setting and query:
<Source>
All waveforms that can be displayed, see
Usage:
Asynchronous command
DISPlay:SIGNal:LABel:FONTsize
<Fontsize>
Sets the font size of waveform labels.
Parameters:
<Fontsize>
Range:
7 to 30
Increment: 1
*RST:
15
Usage:
Asynchronous command
DISPlay:SIGNal:LABel:ADD
<LabelID>, <Source>, <LabelText>, <PositionMode>,
<XPositon>, <YPositon>
Creates a new waveform label for the specified source waveform.
Setting parameters:
<LabelID>
String with the label identifier. The <LabelID> and <Source>
parameters identify each label uniquely, so the label ID must be
unique for the given waveform. Note the <LabelID> because it is
not possible to query it, or to read it on the user interface.
Instrument setup