Protocol analysis
R&S
®
RTO6
896
User Manual 1801.6687.02 ─ 05
1. Press [Bus] on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Setup" tab, select the protocol: M-PHY.
4. Tap "Variant" and select the protocol.
5. Enter the number of "Lanes".
6. Select the source "Channel" for each lane.
7. Enter the "Thresholds" for each lane.
8. Enable "Decode".
13.24.3
M-PHY trigger
13.24.3.1
M-PHY trigger settings
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
Make sure that:
●
In the [Bus] > "Setup" tab:
–
The correct "Protocol" is selected.
–
The data sources of the serial bus are channel signals.
●
In the "Trigger" > "Setup" tab:
–
The "Trigger on" = "Single event"
–
The trigger "Source" = "Serial bus".
–
The correct "Serial bus" is selected.
Type
Selects the trigger type for the analysis of the selected protocol.
M-PHY and USB SSIC (option R&S
RTO6-K580 and K570)