Remote control commands
R&S
®
RTO6
1786
User Manual 1801.6687.02 ─ 05
<n>
*
Parameters:
<BitState>
ONE | ZERO | DC
ONE
1
ZERO
0
DC
"Do not care" (DC) = X
*RST:
DC
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CMSB:DMAX
<SearchName>,<Frame>, <Field>, <Data>
SEARch:TRIGger:CMSB:FRAMe<m>:FLD<n>:DMAX
<SearchName>,<DataMax>
SEARch:TRIGger:CMSB:FRAMe<m>:FLD<n>:DMAX?
<SearchName>
Sets the end value of a data pattern range if
is set to
INRange
or
OORange
.
Suffix:
<m>
.
*
<n>
*
Parameters:
<DataMax>
Specifies the name of the generic trigger setting frame.
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CMSB:DMIN
<SearchName>,<Frame>, <Field>, <Data>
SEARch:TRIGger:CMSB:FRAMe<m>:FLD<n>:DMIN
<SearchName>,<DataMin>
SEARch:TRIGger:CMSB:FRAMe<m>:FLD<n>:DMIN?
<SearchName>
Specifies the data pattern to be searched, or sets the start value of a data pattern
range to be searched.
Suffix:
<m>
.
*
<n>
*
Parameters:
<DataMin>
Specifies the name of the generic trigger setting frame.
Parameters for setting and query:
<SearchName>
Protocols