Pulse source and TDR/TDT analysis (options R&S
RTO6-B7/K130)
R&S
®
RTO6
1045
User Manual 1801.6687.02 ─ 05
When fastening the cable, observe the maximum torque of 0.9
Nm. Therefore, using a
0.9
Nm torque wrench is recommended (optional accessory 1328.8534.35).
To ensure accurate and reliable measurements, connect only cables with a maximum
length of 3
m.
Always use double-shielded cables to prevent electromagentic interference (EMI) from
impairing measurement results.
Always use cables to connect the DUT or test fixture to the pulse source outputs.
Never connect the DUT or test fixture directly to the outputs.
16.1.1
Settings for the pulse source
Access: [App Cockpit] > "Generators" > "PSC"
Terminate all unused outputs of the pulse source with 50
Ω
terminating resistors.
Enable
Switches all four pulse source outputs on and off.
Remote command:
Pulse source (option R&S
RTO6-B7)