Acquisition and waveform setup
R&S
®
RTO6
149
User Manual 1801.6687.02 ─ 05
"Find trigger
level"
Sets the trigger level automatically to
0.5 * (MaxPeak – MinPeak)
.
"Create report" Creates and saves a report using the report settings. See also
"Probe mode"
Only available if a R&S
RT
‑
ZM modular probe is connected. Sets the
measurement mode of the modular probe. See also
"Probe Setup"
Opens the "Probes Setup" dialog box.
Remote command:
ProbeMeter
The integrated R&S ProbeMeter of active voltage probes with Rohde
&
Schwarz probe
interface is a voltmeter. It measures DC voltages between the probe tip and ground
connection or between the probe tips with very high precision. The R&S ProbeMeter
enables ground-referenced measurements of voltages. The measurement is performed
continuously and in parallel to the measurements of the oscilloscope.
●
"Probemeter"
Select "Probemeter" to activate the integrated R&S ProbeMeter of active R&S
probes. The measured voltages are displayed in the "ProbeMeter" result box on
the screen.
●
ProbeMeter measurement results of single-ended active R&S probes
Measures the voltage between the probe tip and the ground.
●
ProbeMeter measurement results of differential and modular R&S probes
You can select the voltage to be measured by the differential active probe:
–
"Differential / Common Mode":
Differential voltage is the voltage between the positive and negative signal
sockets.
Common mode voltage is the mean voltage between the signal sockets and
the ground socket. It measures the voltage level relative to ground, for exam-
ple, to check the operating voltage window.
–
"Single Ended Pos/Neg": Measures the voltage between the positive/negative
signal socket and the ground.
Probes