Remote control commands
R&S
®
RTO6
1804
User Manual 1801.6687.02 ─ 05
TRIGger<m>:CAN:FDATa:BRS
<BRSBit>
Sets the bit rate switch bit.
Parameters:
<BRSBit>
ONE | ZERO | DC
ONE: the bit rate switches from the bit rate of the arbitration
phase to the faster data rate.
*RST:
ONE
TRIGger<m>:CAN:FDATa:ESI
<ESIBit>
Sets the error state indicator bit.
Parameters:
<ESIBit>
ONE | ZERO | DC
DC: don't care, bit is nor relevant
*RST:
DC
TRIGger<m>:CAN:DCONdition
<DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with
.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with
and
*RST:
EQUal
TRIGger<m>:CAN:DMIN
<DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Protocols