Remote control commands
R&S
®
RTO6
1805
User Manual 1801.6687.02 ─ 05
TRIGger<m>:CAN:DMAX
<DataPatternTo>
Sets the the end value of an data range if
INRange
or
OORange
.
Parameters:
<DataPatternTo>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:CAN:BORDer
<Endianness>
Sets the byte order (endianness) of the data transfer. Only for CAN protocol.
Parameters:
<Endianness>
BENDian | LENDian
BENDian
Big endian, data is analyzed and evaluated in the order of
reception.
LENDian
Little endian, the instrument reads the complete data, reverses
the byte order of the data, and compares it with the specified
data word.
*RST:
BENDian
TRIGger<m>:CAN:DLCCondition
<DLCOperator>
Operator to set the data length code for triggering on CAN and CAN FD data.
For details, see
"Data setup: DLC, NDB, Transfer, Condition, Data min, Data max"
The number of data bytes to be found is set with
.
See also:
Parameters:
<DLCOperator>
EQUal | GETHan
For little endian transfer direction, EQUal must be set.
*RST:
GETHan
TRIGger<m>:CAN:DLC
<WordCount>
Sets the Data Length Code, the number of data bytes to be found. For complete defini-
tion, set also the operator with
Parameters:
<WordCount>
Range:
CAN: 1 to 8, CAN FD: 1 to 15 (64 bytes)
Increment: 1
*RST:
1
Protocols