![R&S RTO6 User Manual Download Page 979](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646979.webp)
Protocol analysis
R&S
®
RTO6
979
User Manual 1801.6687.02 ─ 05
For a description of the color table settings, see
Persistence
Each eye diagram has its specific persistence settings, independent of the general
waveform persistence.
Enable ← Persistence
If enabled, each new data point in the diagram area remains on the screen for the
duration that is defined using
is selected.
If disabled, the waveform points are displayed only for the current acquisition.
Remote command:
EYE<m>:DISPlay:PERSistence:STATe
Infinite persistence ← Persistence
If infinite persistence is enabled, each new waveform point remains on the screen until
this option is disabled. Use infinite persistence to display rare events in the signal.
Remote command:
EYE<m>:DISPlay:PERSistence:INFinite
Time ← Persistence
Sets a time factor that controls how long the waveform points fade away from the dis-
play. Thus, the R&S
RTO6 emulates the persistence of analog phosphor screens.
Remote command:
EYE<m>:DISPlay:PERSistence:TIME
Reset ← Persistence
Resets the display, removing persistent all waveform points.
Remote command:
EYE<m>:DISPlay:PERSistence:RESet
Eye stripe display
Defines when the eye stripe is shown: always, never, or during mask tests.
The eye stripe is shown at the bottom of the main source waveform. It shows the part
of the waveform that contributes to the eye. When a mask test is performed on the eye
diagram, the eye stripe shows the position of the mask violations in red color.
Remote command:
Slices per acquisition
Sets the number of waveform slices for a single acquisition.
Remote command:
DDR (option R&S
RTO6-K91 )