Deembedding (option R&S
RTO6-K121)
R&S
®
RTO6
1204
User Manual 1801.6687.02 ─ 05
Probing, Terminating
Sets the probing mode.
Select "Probing" if you use a high-ohmic probe and measure on an existing line in par-
allel to the load.
Select "Terminating" if the measuring equipment is the load of the line.
Remote command:
DEEMbedding<m>:COMPonent<n>:MODE
Remove probe load
If "Probing" is selected, you can remove the loading of the probe.
If the probe loading is removed, you see the signal that would be at the measurement
point if the probe's input impedance is ideal (infinite impedance).
Without removing the loading, you see the real signal at the measurement point,
including the probe loading.
Remote command:
PROBe<m>:DEEMbedding:REMProbeload
DEEMbedding<m>:COMPonent<n>:LOAD:REMProbeload
Source impedance
Shows the source impedance of the DUT. If an R&S
RT
‑
ZM probe is connected, the
value depends on the selected probe mode: common mode, differential, or single-
ended measurement.
Deembedding settings