Remote control commands
R&S
®
RTO6
1420
User Manual 1801.6687.02 ─ 05
TRIGger<m>:QUALify<n>:AB:LOGic
<Operator>
TRIGger<m>:QUALify<n>:CD:LOGic
<Operator>
TRIGger<m>:QUALify<n>:ABCD:LOGic
<Operator>
Defines the logical combination of the indicated channels after evaluating the previous
logical operations:
●
AB: CH1 and CH2
●
CD: CH3 and CH4
●
ABCD: result of AB and CD
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n>
1..18
Trigger type, see
.
Parameters:
<Operator>
AND | NAND | OR | NOR
AND
logical AND, conjunctive combination
NAND
logical NOT AND
OR
logical OR, disjunctive combination
NOR
logical NOT OR
*RST:
AND
Usage:
Asynchronous command
24.9.16
Holdoff
....................................................................................... 1420
......................................................................................... 1421
......................................................................................1422
........................................................................................... 1422
.......................................................................................... 1423
................................................................................. 1423
.................................................................................... 1423
TRIGger<m>:HOLDoff:MODE
<Mode>
Selects the method to define the holdoff condition.
The trigger holdoff defines when the next trigger after the current will be recognized.
Thus, it affects the next trigger to occur after the current one. Holdoff helps to obtain
stable triggering when the oscilloscope is triggering on undesired events.
Trigger