Remote control commands
R&S
®
RTO6
1423
User Manual 1801.6687.02 ─ 05
TRIGger<m>:HOLDoff:MAX
<RandomMaxTime>
Defines the upper limit for the random time holdoff. The setting is relevant if the holdoff
mode is set to RANDom.
See also:
●
●
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<RandomMaxTime> Range:
100E-9 to 10
Increment: 200E-6
*RST:
2E-3
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:HOLDoff:AUTotime?
Returns the resulting holdoff time if the holdoff mode is set to AUTO:
Auto time = Auto
time scaling * Horizontal scale
. The auto time scaling factor is defined with
.
See also:
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Return values:
<AutoTime>
Holdoff time
Range:
100E-9 to 10
*RST:
1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE AUTO
TRIGger:HOLDoff:SCALing 0.5
TRIGger:HOLDoff:AUTotime?
1ms
Result if the horizontal scale is 1 ns/div
Usage:
Query only
Asynchronous command
TRIGger<m>:HOLDoff:SCALing
<AutoTimeScaling>
Sets the auto time scaling factor the horizontal scale is multipied with:
Auto time = Auto
time scaling * Horizontal scale
. The setting is relevant if the holdoff mode is set to
AUTO.
Trigger