Remote control commands
R&S
®
RTO6
1422
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
TRIGger<m>:HOLDoff:EVENts
<Events>
Defines the number of triggers to be skipped. The next trigger only occurs when this
number of events is reached. The setting is relevant if the holdoff mode is set to
EVENts.
See also:
●
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Events>
Range:
1 to 2147483647
Increment: 1
*RST:
1
Example:
TRIGger:HOLDoff:MODE EVENts
TRIGger:HOLDoff:EVENts 5
Usage:
Asynchronous command
TRIGger<m>:HOLDoff:MIN
<RandomMinTime>
Defines the lower limit for the random time holdoff. The setting is relevant if the holdoff
mode is set to RANDom.
See also:
●
●
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<RandomMinTime>
Range:
100E-9 to 5
Increment: 200E-6
*RST:
1E-3
Default unit: s
Example:
TRIGger:HOLDoff:MODE RANDom
TRIGger:HOLDoff:MIN 1ms
TRIGger:HOLDoff:MAX 2ms
The holdoff time is set randomly between 1 ms and 2 ms.
Usage:
Asynchronous command
Trigger