Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1192
User Manual 1801.6687.02 ─ 05
6. Select the correct channels for the "Current Source" and the "Voltage Source".
7. Select the "Settings" tab.
8. Set the "Trigger" settings according to your signal.
9. Set the "Tube tolerance", "Signal high" and "Signal low" according to your require-
ments.
10. Set the "SMPS switching frequency" according to your device.
11. Select an "Optimize Scaling" option.
12. Tap "Execute".
13. If needed adjust the cursors manually.
On the screen, you can see the measurement of the current and the voltage . Also,
there is a table giving information about important measurement parameters. For
details, see
Chapter 19.6.2.1, "Transient response results"
19.6.2.3
Transient response settings
In the "Channels" tab, you set the current source and the voltage sources, see also:
Chapter 19.2.1, "Channels tab"
Output