Remote control commands
R&S
®
RTO6
1550
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<LimitCheckState>
OFF | LIMit | LMARgin
OFF
No limit check.
LIMit
Only limits are checked.
LMARgin
Limits and margins are checked.
*RST:
OFF
Usage:
Asynchronous command
MEASurement<m>:ONViolation:BEEP
<Beep>
Generates a beep sound for the specified event.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<Beep>
NOACtion | SUCCess | VIOLation
See
Chapter 24.4.5, "Event parameter"
*RST:
NOACtion
Usage:
Asynchronous command
MEASurement<m>:ONViolation:ACQStop
<StopAcq>
Stops data acquisition for the specified event.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<StopAcq>
NOACtion | SUCCess | VIOLation
See
Chapter 24.4.5, "Event parameter"
*RST:
NOACtion
Usage:
Asynchronous command
Automatic measurements