Pulse source and TDR/TDT analysis (options R&S
RTO6-B7/K130)
R&S
®
RTO6
1050
User Manual 1801.6687.02 ─ 05
13. Select "Next".
Now you can analyze the DUT.
16.2.2.2
Measuring the DUT
1. Open [App Cockpit]> "Analysis" > "TDR / TDT"
2. Load an existing calibration file, or calibrate the setup as described in
ter 16.2.2.1, "Calibrating the setup"
3. Discharge the DUT before connecting.
4. Connect the DUT instead of calibration standards as shown in the dialog.
5. Under "Signals", select the characteristics that you want to analyze.
6. Enter the "Bandwidth" of the DUT.
7. Select whether you want to see the time or the distance on the x-axis: "Domain".
8. If the distance is selected, enter the permittivity of the DUT: "
ε
_r".
9. To check the impact of the frequency response calibration on the time domain
waveform, disable "Use frequency response calibration".
10. Close the dialog box.
11. Set the horizontal time scale so that the complete DUT is captured and visible on
the screen.
12. Analyze the selected signals. See also:
Chapter 16.2.6, "Measurement results"
16.2.3
Calibration settings
16.2.3.1
Setup: TDR/TDT configuration
Access: [App Cockpit]> "Analysis" > "TDR / TDT"
TDR/TDT analysis (option R&S
RTO6-K130)