Protocol analysis
R&S
®
RTO6
766
User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:RFFE:ERRor:NOResponse
TRIGger<m>:RFFE:ERRor:USEQuence
Slave Addr. /MID
Defines the address of the slave / master ID. The slave address setup consists of the
condition and one or two address patterns.
"Condition"
Sets the operator to trigger on a specific address pattern ("Equal" or
"Not equal") or an address range.
"Slave Address
(Min)"
Defines the slave address pattern for all operators that require one
pattern.
"Slave Adress
(Max)"
Defines the second address pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:SADD:CONDition
Address
Defines the register address. The register address setup consists of the condition and
one or two address patterns.
"Condition"
Sets the operator to trigger on a specific address pattern ("Equal" or
"Not equal") or an address range.
"Address
(Min)"
Defines the register address pattern for all operators that require one
pattern.
"Adress (Max)" Defines the second address pattern that is required to specify a range
with conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:RFFE:ADDRess:CONDition
Byte count
Sets the number of data frames to be read or written in the command sequence. The
setting is available for all "Extended Register" command sequences, which can trans-
fer more than one data frame.
"Condition"
Sets the operator to trigger on a specific byte count ("Equal" or "Not
equal") or a byte count range.
"Byte/Bit count
(Min)"
Defines the byte count for all operators that require one count setting.
RFFE (option R&S
RTO6-K550)