Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1124
User Manual 1801.6687.02 ─ 05
5. Select the correct channels for the "Current Source" and the "Voltage Source".
6. Tap "Next".
7. Check and complete the probe setup in the "Probes" tab.
Current probes and high-voltage differential probes are not automatically detected
by the instrument. Tap "Predefined probe" and select the correct probe type.
8. Tap the "Settings" tab.
9. Set the
"Overwrite present skew setup"
on page 1127 options. These settings define whether the instrument uses
the deskew result for user-defined preset and general skew settings.
10. Tap "Auto deskew".
The probes are deskewed and the measurement can be started.
If no deskew fixture is available, you have to deskew your probes manually, see
Probes
In the "Probes" tab, you check and set up your voltage and current probes.
Power measurement selection