Make sure that:
●
In the [Bus] > "Setup" tab:
–
The correct "Protocol" is selected.
–
The data sources of the serial bus are channel signals.
●
In the "Trigger" > "Setup" tab:
–
The "Trigger on" = "Single event"
–
The trigger "Source" = "Serial bus".
–
The correct "Serial bus" is selected.
Frame Type
Selects the frame type for the USBPD analysis. For the available packet types, see
You can define individual checking parameters for the fields listed of the frame.
To specify these parameters, select a field from this list and open "Set details". You can
define the data and/or index operators and values, or the bit state.
The trigger instant is the last criterion that is fulfilled.
Remote command:
TRIGger<m>:USBPd:FRAMe<n>:ENABle