![R&S RTO6 User Manual Download Page 694](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_3064646694.webp)
Protocol analysis
R&S
®
RTO6
694
User Manual 1801.6687.02 ─ 05
"Condition"
Sets the comparison condition to a specific value or a range.
"RTA Min/
RTA"
Specifies the value or sets the start value of a range.
"RTA Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:MILStd:CDST:RCONdition
TRIGger<m>:MILStd:CMD:RCONdition
TRIGger<m>:MILStd:DATA:RCONdition
Data Pattern
Sets the data to be triggered on. The trigger data setup consists of a comparison con-
dition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:MILStd:DATA:DCONdition
Data Index
The MIL-MIL-STD-1553 standard defines the length of a message to a series of up to
32 words. Data index sets the range within this series of the data words that is consid-
ered for the analysis. The data index setup consists of the condition and one or two
data index values.
"Condition"
Sets the comparison condition to a specific value or a range.
"Index Min/Index"
Specifies the value or sets the start value of a range.
"Index Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
MIL-1553 (option R&S
RTO6-K530)