Protocol analysis
R&S
®
RTO6
903
User Manual 1801.6687.02 ─ 05
"Reserved"
Reserved bit command. These bits are reserved for future use.
Remote command:
WORD 1/WORD 2/WORD 3/WORD 4
Sets the specified words to be triggered on. The setup for a word consists of the condi-
tion and one or two word patterns.
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/Data"
Specifies the value or sets the start value of a range.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
For " WORD 1":
TRIGger<m>:MPHY:LWONe:CONDition
For " WORD 2":
TRIGger<m>:MPHY:LWTWo:CONDition
For " WORD 3":
TRIGger<m>:MPHY:LWTHree:CONDition
For " WORD 4":
TRIGger<m>:MPHY:LWFour:CONDition
TC
Sets the specified Traffic Class (TC) to be triggered on.
TC is used for classifying the traffic flow based on protocol and port number, e.g. TC0
and TC1 are two priority classes with guaranteed link reliability defined and used in L2.
"Data"
Defines the bit pattern of the TC pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
Remote command:
M-PHY and USB SSIC (option R&S
RTO6-K580 and K570)