Network operation and remote control
R&S
®
RTO6
1250
User Manual 1801.6687.02 ─ 05
Bit No.
Meaning
8
Overload on external trigger input
9
Overload warning for external trigger input
STATus:QUEStionable:TEMPerature register
This register contains information about the instrument's temperature.
Table 22-5: Bits in the STATus:QUEStionable:TEMPerature register
Bit No.
Meaning
0
TEMP WARN
This bit is set if a temperature warning on channel 1, 2, 3 or 4 occurred.
1
TEMP ERRor
This bit is set if a temperature error on channel 1, 2, 3 or 4 occurred.
STATus:QUEStionable:ADCState register
This register contains all information about overflow of the ADC.
The bit is set if the assigned channel signal is clipped on the upper or lower edge of the
screen. In this case, the signal does not fit in the range of the ADC and overflow
occurs.
Table 22-6: Bits in the STATus:QUEStionable:ADCState register
Bit No.
Meaning
0
CHANnel1
, clipping on the upper limit
1
CHANnel1
, clipping on the lower limit
2
CHANnel2
, clipping on the upper limit
3
CHANnel2
, clipping on the lower limit
4
CHANnel3
, clipping on the upper limit
5
CHANnel3
, clipping on the lower limit
6
CHANnel4
, clipping on the upper limit
7
CHANnel4
, clipping on the lower limit
STATus:QUEStionable:FREQuency register
Table 22-7: Bits in the STATus:QUEStionable:FREQuency register
Bit No.
Meaning
0
Oven cold. This bit is set if the reference oscillator has not yet attained its operating tempera-
ture.
Remote control