Pulse source and TDR/TDT analysis (options R&S
RTO6-B7/K130)
R&S
®
RTO6
1058
User Manual 1801.6687.02 ─ 05
See also:
Chapter 12.2, "Save and recall waveform data and results"
Remote commands:
●
●
●
16.2.7
Troubleshooting
If the calibration or measurement does not show meaningful results, the following table
helps you to find the cause.
Issue
Checks
Calibration
No waveform is visible.
Check the cabling.
Check the pulse source.
Calibration
Calibration with short failed
Check the cable length: 0.5
m
<
length
<
3
m
Use cables with low loss.
Use a short that has a low impedance for high frequen-
cies.
Calibration
Calibration with open failed
Use cables with low loss.
Calibration
Calibration with match failed
Use cables with low loss.
Use match with 50
Ω±
10% for single-ended measure-
ments.
Calibration
Calibration with through failed
Use cables with low loss.
Check the cable length: 0.5
m
<
length
<
3
m
Calibration
Failure after completion of all
calibration steps
Use cables with low loss.
Check the pulse source.
Measurement
Waveform is noisy or unstable Evaluate the calibration results for any resonances or
high attenuation in frequency response.
Increase averaging.
Reduce bandwidth.
Disable calibration. If noise disappears, reduce attenua-
tion or remove resonances from your setup and recali-
brate.
TDR/TDT analysis (option R&S
RTO6-K130)