Remote control commands
R&S
®
RTO6
1710
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
BUS<m>:SPI:SSELect:POLarity
<SSPolarity>
Selects whether transmitted slave select signal is high active (high = 1) or low active
(low = 1).
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SSPolarity>
ACTLow | ACTHigh
*RST:
ACTLow
Usage:
Asynchronous command
BUS<m>:SPI:MISO:SOURce
<MISOSource>
Sets the input channel of the MISO line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MISOSource>
NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See
Chapter 24.4.2, "Waveform parameter"
*RST:
None
Usage:
Asynchronous command
BUS<m>:SPI:MISO:POLarity
<MISOPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1).
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MISOPolarity>
ACTLow | ACTHigh
*RST:
ACTHigh
Usage:
Asynchronous command
Protocols