![R&S RTO6 User Manual Download Page 2312](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462312.webp)
Remote control commands
R&S
®
RTO6
2312
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:MPHY:PBEGin:CONDition
<SearchName>,<CondPACPBegin>
SEARch:TRIGger:MPHY:PBEGin:CONDition?
<SearchName>
Sets the condition for PACP Begin. You can define an exact value or a value range.
Parameters:
<CondPACPBegin>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with the corresponding
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with
SEARch:TRIGger:MPHY:PBEGin:MIN
and
SEARch:TRIGger:MPHY:PBEGin:MAX
*RST:
EQUal
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:PBEGin:MIN
<SearchName>,<PACPBEGINMin>
SEARch:TRIGger:MPHY:PBEGin:MIN?
<SearchName>
Specifies the PACP Begin bit pattern, or sets the start value of a pattern range.
Parameters:
<PACPBEGINMin>
Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:MPHY:PBEGin:MAX
<SearchName>,<PACPBEGINMax>
SEARch:TRIGger:MPHY:PBEGin:MAX?
<SearchName>
Sets the end value of the PACP Begin range if
is set to
INRange
or
OORange
.
Parameters:
<PACPBEGINMax> Numeric or string pattern, see
on page 1286. The string parameter accepts the bit
value X (don't care).
Parameters for setting and query:
<SearchName>
Protocols