Remote control commands
R&S
®
RTO6
1345
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Name>
Name string
Usage:
Query only
Asynchronous command
PROBe<m>:SETup:BANDwidth?
Queries the bandwidth of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<Bandwidth>
Range:
10000 to 20E+9
*RST:
1E+9
Default unit: Hz
Usage:
Query only
Asynchronous command
PROBe<m>:SETup:ATTenuation[:AUTO]?
Queries the attenuation of the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Return values:
<PrbAttMdAuto>
Range:
1E-3 to 1000
*RST:
1
Default unit: V/V
Usage:
Query only
Asynchronous command
PROBe<m>:SETup:OFFSet:AZERo
Performs an automatic correction of the zero error. If the DUT is ground-referenced,
the AutoZero function can improve the measurement results.
See also:
"Detect AutoZero, Use AutoZero"
Suffix:
<m>
.
1..4
Selects the input channel.
Usage:
Event
Asynchronous command
Acquisition and setup