Acquisition and waveform setup
R&S
®
RTO6
175
User Manual 1801.6687.02 ─ 05
Measurement Impedance
Sets the impedance of the probe channel for power calculations and measurements.
Remote command:
5.7.1.3
Zero compensation
Access: [App Cockpit]> "Analysis" > "Multi-channel probe"> "ZeroComp" tab
Zero compensation is used to avoid negative currents in measurement results, and to
improve the measurement accuracy. The determined compensation is valid as long as
the temperature is constant and the probe settings are unchanged.
1. Disconnect the DUT from the power supply.
2. Adjust the current settings, in particular, the shunt settings. See
3. Connect one probe tip of the R&S
RT-ZVC probe to the high voltage pin of the
DUT. Disconnect the other probe tip from the DUT.
4. Open the "ZeroComp" settings: [App Cockpit]> "Analysis" > "Multi-channel probe">
"ZeroComp" tab
5. Tap "ZeroComp" to determine the zero offset.
6. Enable "Use" to compensate for the zero offset.
ZeroComp
Measures the zero offset, the mean value on a currentless DUT. If temperature
changes, or if you change the shunt or other probe settings, repeat the measurement.
Remote command:
Use
If enabled, the measured "ZeroComp" offset value is compensated automatically.
Remote command:
R&S
RT-ZVC probe