Waveform generator (Option R&S
RTO6-B6)
R&S
®
RTO6
1027
User Manual 1801.6687.02 ─ 05
Start frequency
Sets the start frequency of the sweep signal.
Remote command:
Stop freuqency
Sets the stop frequency of the sweep signal.
Remote command:
Sweep time
Sets the duration of the sweep.
Remote command:
15.1.5
Arbitrary
The arbitrary waveform generator allows you to output a user-defined waveform for
testing your devices. You can output a waveform from a file or from the current
R&S
RTO6 reference curve format. Files in
*.csv
and
*.bin
formats are supported.
This files must follow a defined structure. You can load
*.csv
files in an R&S Wave
Setup of the waveform generator