Protocol analysis
R&S
®
RTO6
759
User Manual 1801.6687.02 ─ 05
BUS<m>:RFFE:DATA:THReshold:HIGH
BUS<m>:RFFE:DATA:THReshold:LOW
BUS<m>:RFFE:DATA:THReshold:HYSTeresis
Show thresholds
If enabled, the threshold levels are displayed in the diagram.
Remote command:
13.17.2.2
Advanced settings
Access: [Bus] > "Advanced" tab.
Autoset
Starts software algorithms for determining the signal threshold levels and bitrate.
Remote command:
Glitch filter
Enables the glitch filter on the SCLK and SDATA lines to improve decode accuracy.
The "Glitch filter width" field sets the maximum glitch width to be ignored.
Remote command:
Timing
The gap defines the idle time between the Bus Park Cycle (BP) and Sequence Start
Condition (SSC).
You can define a "Min gap time". The standard defines a minimum of 10 ns to separate
two subsequent command sequences.
RFFE (option R&S
RTO6-K550)