Remote control commands
R&S
®
RTO6
1838
User Manual 1801.6687.02 ─ 05
BUS<m>:LIN:TECHnology
<Technology>
Sets the threshold voltage as defined for various signal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology>
V15 | V25 | V35 | V6 | V9 | MANual
V15
1.5 Volt (TTL)
V25
2.5 Volt (CMOS 5.0 V)
V35 | V6 | V9
3.5 V (7 V supply), 6.0 V (12 V supply), 9.0 V (18 V supply)
respectively
MANual
Manual setting of user-defined values with
*RST:
V35
BUS<m>:LIN:BITRate
<Bitrate>
Sets the number of transmitted bits per second.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Bitrate>
Range:
1000 to 20000
Increment: 1
*RST:
9600
Default unit: bps
BUS<m>:LIN:POLarity
<Polarity>
Defines the idle state of the bus. The idle state is the rezessive state and corresponds
to a logic 1.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Polarity>
IDLLow | IDLHigh
*RST:
IDLHigh
Protocols