Remote control commands
R&S
®
RTO6
2413
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
EYE<m>:QUALify:SIGNal:SOURce
<Source>
Selects the waveform to be compared.
Suffix:
<m>
.
1..4
Parameters:
<Source>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
Usage:
Asynchronous command
EYE<m>:QUALify:SIGNal:CONDition
<Condition>
Selects the comparison condition for the the waveform level.
Suffix:
<m>
.
1..4
Parameters:
<Condition>
LTHan | GTHan | INRange | OORange
*RST:
GTHan
Usage:
Asynchronous command
EYE<m>:QUALify:SIGNal:LSELect
<LevelSelection>
Selects the threshold. You can use the lower, middle, or upper reference level. These
levels are the reference levels that are defined for measurements.
Suffix:
<m>
.
1..4
Parameters:
<LevelSelection>
UPPer | MIDDle | LOWer
*RST:
MIDDle
Example:
The slices of the channel 1 signal only contribute to the eye gen-
eration if the voltage level of the signal on channel 2 is higher
than the upper reference level.
EYE:SOURce C1W1
EYE:QUALify:SIGNal:SOURce C2W1
EYE:QUALify:SIGNal:CONDition GTHan
EYE:QUALify:SIGNal:LSELect UPPer
EYE:QUALify:SIGNal ON
Usage:
Asynchronous command
Protocols