Remote control commands
R&S
®
RTO6
1351
User Manual 1801.6687.02 ─ 05
PROBe<m>:SETup:ATTenuation:MODE
<PrbAttMd>
Set the mode to MANual if the instrument does not detect the probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<PrbAttMd>
AUTO | MANual
*RST:
AUTO
Usage:
Asynchronous command
PROBe<m>:SETup:ATTenuation:UNIT
<AttenuationUnit>
Sets the unit for the connected probe type if
PROBe<m>:SETup:ATTenuation:MODE
MANual
.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<AttenuationUnit>
V | A | W
Voltage probe (V), current probe (A), power probe (W)
*RST:
V
Usage:
Asynchronous command
PROBe<m>:SETup:ATTenuation:MANual
<ManualAttenuation>
Sets the attenuation for the connected probe if
MANual
.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualAttenuation> Range:
100E-6 to 10000
Increment: 0.1
*RST:
1
Default unit: Depends on the selected unit
Usage:
Asynchronous command
PROBe<m>:SETup:GAIN:MANual
<GainManual>
Sets the gain of a current probe.
Acquisition and setup