Remote control commands
R&S
®
RTO6
1403
User Manual 1801.6687.02 ─ 05
Parameters:
<WidthDelta>
Range:
0 to 10
Increment: 100E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
24.9.9
Slew rate trigger
.......................................................................................... 1403
TRIGger<m>:LEVel<n>:SLEW:LOWer
........................................................................... 1403
TRIGger<m>:LEVel<n>:SLEW:UPPer
............................................................................1403
......................................................................................... 1404
............................................................................................ 1404
...........................................................................................1405
TRIGger<m>:SLEW:SLOPe
<Slope>
Selects the edge type for the trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope>
POSitive | NEGative | EITHer
See
Chapter 24.4.3, "Slope parameter"
*RST:
POSitive
Usage:
Asynchronous command
TRIGger<m>:LEVel<n>:SLEW:LOWer
<Level>
TRIGger<m>:LEVel<n>:SLEW:UPPer
<Level>
Set the lower and upper voltage thresholds. When the signal crosses a level, the slew
rate measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n>
1..31
Indicates the trigger source:
1...4 = channel 1...4
5...27 = not available
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Trigger