Remote control commands
R&S
®
RTO6
1405
User Manual 1801.6687.02 ─ 05
Parameters:
<Time>
Range:
100E-12 to 10000
Increment: 100E-9
*RST:
100E-12
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:SLEW:DELTa
<TimeDelta>
Defines a time range around the slew rate specified using
.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<TimeDelta>
Range:
0 to 10
Increment: 100E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
24.9.10
Setup & Hold
The Setup & Hold trigger is only available for the A-event (Suffix = 1).
TRIGger<m>:SETHold:CSOurce:EDGE
......................................................................... 1405
TRIGger<m>:SETHold:CSOurce:LEVel
..........................................................................1406
TRIGger<m>:SETHold:CSOurce[:VALue]
....................................................................... 1406
.......................................................................................1406
....................................................................................... 1407
...............................................................................................1407
TRIGger<m>:SETHold:CSOurce:EDGE
<ClockEdge>
Sets the edge of the clock signal to define the time reference point for the setup and
hold time.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockEdge>
POSitive | NEGative | EITHer
See
Chapter 24.4.3, "Slope parameter"
*RST:
POSitive
Usage:
Asynchronous command
Trigger