TRIGger<m>:SETHold:CSOurce:LEVel
<ClockLevel>
Sets the voltage level for the clock signal. Both this command and
define the starting point for calculation of the setup and
hold time.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockLevel>
Range:
-10 to 10
Increment: 1E-3
*RST:
0
Default unit: V
Usage:
Asynchronous command
TRIGger<m>:SETHold:CSOurce[:VALue]
<ClockSource>
Selects the source of the clock signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<ClockSource>
CHAN1 | CHANnel1 | CHAN2 | CHANnel2 | CHAN3 |
CHANnel3 | CHAN4 | CHANnel4
CHAN1 = CHANnel1, CHAN2 = CHANnel2, CHAN3 = CHAN-
nel3, CHAN4 = CHANnel4
Input channel of the clock signal
*RST:
CHAN1
Usage:
Asynchronous command
TRIGger<m>:SETHold:HTIMe
<HoldTime>
Sets the minimum time
after
the clock edge while the data signal must stay steady
above or below the data level.
The hold time can be negative. In this case, the setup time is always positive. The
setup/hold interval starts before the clock edge (setup time) and ends before the clock
edge (hold time). If you change the negative hold time, the setup time is adjusted by
the intrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Trigger