Remote control commands
R&S
®
RTO6
1407
User Manual 1801.6687.02 ─ 05
Parameters:
<HoldTime>
Range:
-99.999E-9 to 100E-9
Increment: 1E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:SETHold:STIMe
<SetupTime>
Sets the minimum time
before
the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. The
setup/hold interval starts after the clock edge (setup time) and ends after the clock
edge (hold time). If you change the negative setup time, the hold time is adjusted by
the instrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SetupTime>
Range:
-99.999E-9 to 100E-9
Increment: 1E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
TRIGger<m>:SCOupling
<TrigLevSrcCpl>
Sets the trigger levels of all used channels to the level of the data line (Setup & Hold
and serial pattern trigger), clock line (state trigger) or source line (pattern trigger).
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<TrigLevSrcCpl>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
24.9.11
State trigger
The state trigger combines the edge trigger settings with trigger qualification. It is only
available for the A-event (Suffix = 1).
Use the following commands:
●
Trigger