![R&S RTO6 User Manual Download Page 1370](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461370.webp)
Remote control commands
R&S
®
RTO6
1370
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<VertPosi>
Range:
-5 to 5
Increment: 0.02
*RST:
0
Default unit: div
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:SCALe
<VerticalScale>
Sets the vertical scale for the current channel in Volts per division. Increasing the scale
compresses the display of the signal.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<VerticalScale>
Range:
1E-15 to 10E+24
Increment: 10E-6
*RST:
3
Default unit: V/div
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:EVALue
<ExtShuntVal>
Defines the value of the external shunt resistor to calculate the correct current values.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<ExtShuntVal>
Range:
10E-6 to 10000
Increment: 0.02
*RST:
1
Default unit: Ω
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:SHUNt:MODE
<ShuntMode>
Sets the internal or external shunt mode.
Suffix:
<m>
.
1..2
Acquisition and setup