![R&S RTO6 User Manual Download Page 1184](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461184.webp)
Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1184
User Manual 1801.6687.02 ─ 05
The measurement also includes the AC-RMS of the output DC signal, that is calculated
as a standard derivation.
Required probes:
●
Voltage probe
●
(Optional) Current probe
19.6.1.1
Output ripple results
The results of "Output Ripple" measurements are provided in two ways:
●
The diagram shows the graphical presentation of:
–
the voltage waveform
–
(optional) the current waveform
●
The result box displays the numeric measurement results for the voltage and for
the current ripple. For details, see
and
To measure and display the output ripple, the instrument uses the following measure-
ments and waveforms:
●
"MG 7" to measure the current
●
"MG 8" to measure the voltage
The used resources are listed in the "Details" tab. See also:
Table 19-5: Properties output ripple
Mea
s.
type
Symbol
Description/Result
Max
X
Max
Absolute maximum value of the waveform.
Min
X
Min
Absolute minimum value of the waveform.
Output