Remote control commands
R&S
®
RTO6
1353
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
PROBe<m>:SETup:ACCoupling
<ProbeCouplingAC>
Enables AC coupling in the R&S
RT
‑
ZPR20 probe, which removes DC and very low-
frequency components. The R&S
RT
‑
ZPR20 probe requires 50
Ω
input termination,
for which the channel AC coupling is not available. The probe setting allows AC cou-
pling also at 50
Ω
inputs.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeCouplingAC> ON | OFF
*RST:
OFF
Usage:
Asynchronous command
PROBe<m>:SETup:ADVanced:PMToffset
Sets the measured ProbeMeter value as offset. Make sure that the ProbeMeter is
active before you use this command.
Only available for power rail probes R&S
RT
‑
ZPR.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage:
Event
Asynchronous command
PROBe<m>:SETup:ADVanced:RANGe
<ProbeRange>
Sets the voltage range of a R&S
RT-ZHD probe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeRange>
AUTO | MHIGh | MLOW
AUTO
.
MHIGh
Sets the higher voltage range of the connected probe. To query
the value, use
PROBe<m>:SETup:ATTenuation[:AUTO]?
MLOW
Sets the lower voltage range of the connected probe. To query
the value, use
PROBe<m>:SETup:ATTenuation[:AUTO]?
Acquisition and setup