Network operation and remote control
R&S
®
RTO6
1245
User Manual 1801.6687.02 ─ 05
On the touch screen, two buttons appear in the upper left corner: "Local" and "View".
22.5.3.2
Using the display during remote control
You can observe the screen while a remote control script is executed. This is helpful for
program test purposes but tends to slow down the measurement. Therefore it is rec-
ommended that you switch off the display in real measurement applications where a
tested program script is to be executed repeatedly.
► To switch on the display, do one of the following:
● Tap the "View" button in the upper left corner of the touch screen.
● Use the
SYSTem:DISPlay:UPDate ON
command.
► To switch off the display, do one of the following:
● Tap the "View" button again.
● Use the
SYSTem:DISPlay:UPDate OFF
command.
22.5.3.3
Returning to manual operation
The instrument switches back to manual operation when the remote connection is
closed. Besides, you can return to manual operation manually or via remote control.
► To return to manual operation:
● Tap the "Local" button in the upper left corner of the touch screen.
● VXI-11 protocol: Use
>L
interface message.
22.5.4
Remote control - status reporting system
The status reporting system stores all information on the current operating state of the
instrument, and on errors which have occurred. This information is stored in the status
registers and in the error queue. Both can be queried via
STATus
... commands.
22.5.4.1
Hierarchy of status registers
As shown in the following figure, the status information is of hierarchical structure.
Remote control