Remote control commands
R&S
®
RTO6
2349
User Manual 1801.6687.02 ─ 05
Parameters:
<ThresholdL3High>
Range:
-5 to 5
Increment: 1E-3
*RST:
0.02
Default unit: V
BUS<m>:PCIE:LZER:THRLow
<ThresholdL0Low>
BUS<m>:PCIE:LONE:THRLow
<ThresholdL1Low>
BUS<m>:PCIE:LTWO:THRLow
<ThresholdL2Low>
BUS<m>:PCIE:LTHRee:THRLow
<ThresholdL3Low>
BUS<m>:PCIE:SMLanes
<Active>
Enables the display of the result in multiple lanes.
Suffix:
<m>
.
1..4
Parameters:
<Active>
ON | OFF
*RST:
ON
24.17.27.2
Clock data recovery
.......................................................................................... 2349
........................................................................................2350
............................................................................................2350
......................................................................................2350
BUS<m>:CDR:BITRate
<Bitrate>
Sets the quiescent frequency of the PLL. It corresponds to the data rate of the data
stream from which the clock is to be recovered.
Suffix:
<m>
.
1..4
Parameters:
<Bitrate>
Range:
100 to 5E+9
Increment: 10
*RST:
1E+9
Default unit: bps
BUS<m>:CDR:PLL:BWIDth
<PLLBandwidth>
Sets the PLL bandwidth. It defines the part of the spectrum that the PLL can follow dur-
ing synchronization. The PLL bandwidth is usually defined by the transmission stan-
dard.
Protocols